6 in 1 benchtop water test analyzer
6 in 1 benchtop water test analyzer
The PH-2603 multi-parameter meter has been specially designed for the maximum facility of use. The handy key board for the selection of the various functions. The possibility of automatically correcting errors in measurement due to temperature .The display on the front panel reveals the PH mV EC CF or TDS readings taken by the instrument. This instrument has the battery charge function, you will be able to carry out field measurement. Moreover, PH-2603 features a low battery indicator that alerts the user when the battery needs recharging.
6 in 1 benchtop water test analyzer
MODEL | PH-2603
|
Measuring range | pH: 0.00~14.00 mV: 0~ ±1999mV °C: 0~100°C EC: 0.00~19.99ms/cm CF: 0.0~199.9CF TDS: 10~19990ppm |
Resolution | pH: 0.01pH mV: 1mV °C: 0.1°C EC: 0.01ms/cm CF: 0.1CF TDS: 10ppm |
Accuracy | pH: ±0.01pH mV: ±1mV °C: ±0.4°C EC.CF.TDS: ±2%F.S |
Input impedance | 102 Ω |
Calibration | 1 or 2point |
Working temperature | 0~95°C |
Power | DC9V/200mA or Rechargeable battery |
Operating conditions | 5~60°C,RH≤95%(Without magnetic interference around the meter & conspicuous centrum) |
Dimension | 208×200×76mm |
Weight | 510g |
6 in 1 benchtop water test analyzer
PH MEASUREMENTS: Turns on the PH electrode the temperature sensor and the power source. The DC/RB switch must choose DC Turn the meter ON by pressing the ON/OFF key. Remove the PH electrode protective cap, then submerge the PH electrode and the temperature probe into the sample to be tested. Stir gently and wait for the stability symbol. The display will show the PH value automatically compensated for temperature. After use, rinse the electrode with clean water, replace the electrode protective cap.
6 in 1 benchtop water test analyzer
ORP MEASUREMENTS: Oxidation Reduction Potential(ORP) measurements provide a quantification of the oxidizing or reducing power of the sample tested. Connect the ORP electrode to the PH connector. To enter the mV mode turn the instrument ON and press the mV key until the display changes to mV. Then submerge the ORP electrode into the sample to be tested and allow time for the reading to stabilize.
6 in 1 benchtop water test analyzer
TEMPERATURE MEASUREMENTS:Make sure the temperature probe is connected to the meter. Turn the instrument ON and press the key to select the temperature mode. Dip the temperature probe into the sample, allow the reading to stabilize and read the temperature value.
6 in 1 benchtop water test analyzer
EC CF or TDS MEASUREMENTS:Connect the EC electrode to the EC connector. To enter the EC CF or TDS mode turn the instrument ON and press the MODE key until the display changes to ms/cm CF or ppt. Then submerge the EC electrode into the sample to be tested and allow time for the reading to stabilize.