This versatile controller is ideal for many applications, such as ORP monitoring bleaching processes, waste water treatment and swimming pools. ORP-2626 permits automatic control of installations previously checked manually.
The instrument orp water tester can be set for reducing or oxidizing dosage. It measures in the ±1999 mV range, with 1 mV resolution. Set point adjustment and calibration are simply performed through trimmers on the front panel.
The overtime control system advises users when the relay is active too long, helping to prevent over dosage.
Feature
• Easy to handle
• Common BNC connector
• -500-500mV set range
• ORP and temperature two channel display
• Temperature display in °F and °C
Technical data
Items | ORP | Temperature |
Measuring range | -1999~1999mV | -9.9~99.9°C(16~210°F) |
Resolution | 1mV | 0.1°C(0.1°F(16~99°F) |
1°F(100~ 210°F)) | ||
accuracy | ±5mV | ±1°C(±2°F) |
Calibration | 1point or 2point | |
Set range | -500~500mV | |
Standard of CPS | 5A/AC 250V | |
Power | AC230V± 10%, 50HZ±2% | |
Operating conditions | 0~60°C, RH<95% | |
ATC | 0~60°C | |
Revise | 1 point | |
Dimension | 228×163×78mm | |
Weight | 774g |
About ORP
ORP: Redox potential (also known as oxidation / reduction potential, ORP, pe, ε, or Eh) is a measure of the tendency of a chemical species to acquire electrons from or lose electrons to an electrode and thereby be reduced or oxidized, respectively. Redox potential is measured in volts (V), or millivolts (mV). Each species has its own intrinsic redox potential;